[Audio] BIO GENE PRESENTS BIO SAFETY CABINET MODEL: BTI - BSC 3 2 2 BIO TECHNOLOGIES INC.

Page 2 (16s)

[Audio] APPPLICATIONS Bio Gene Bio Safety Cabinet is widely used in Research Centers, Major Universities, Pharmacy, Biotechnology, Medical research, Hospital, Diagnostics, Agriculture science, Pharmacology, Immunology, Genetics, Bioengineering Biotechnology, Microbiology, Botany, Zoology etc. BIO SAFETY CABINET.

Page 3 (52s)

[Audio] WORK SURFACE IS MADE UP OF STAINLESS STEEL 304 TOUGHENED GLASS DOOR HAVING THICKNESS OF 254MM. SWITCH FOR FLUROSCENT LIGHT NOZZLE BIO SAFETY CABINET INTERIOR MICRO PROCESSOR CONTROLLER BASED WITH LCD DISPLAY, MENU CONTROL, AUTOMATIC FILTER ALARM SYSTEM, HAVE THE FOLLOWING MENU CONTROLLED.

Page 4 (1m 29s)

[Audio] HMI CONTROLLER 1. 16 x 2 LCD Alphanumeric display showing the controlling of the different parameters with soft touch keys to control the different parameters. 2. Timers for U.V light operation up to 99mins 3. Timer for Blower operation up to 99:99 hrs 4. Alarm indication on timer completion process. 5. LED indication for all outputs..

Page 5 (2m 7s)

[Audio] UP button is Used for Up Counter for all type of setting DOWN BUTTON is Used for Down Counter for all type of setting MENU button open the Menu & edit the settings UV LIGHT- TURN ON/OFF. FAN - TURN ON/OFF the FAN. POWER BUTTON is used To turn ON/OFF the system. LIGHT button is used To turn ON/OFF the light. ESC( escape) BUTTON is used To escape from all the operation.

Page 6 (2m 43s)

[Audio] Airflow System:  Biological Safety Cabinet is designed with inward air flow at a velocity to protect personnel, HEPA-filtered vertical laminar airflow for product protection, and HEPA-filtered exhausted air for environmental protection  30% air recirculation  70% air exhaust  Maintain a minimum average inflow velocity is 0.65 m/s and average down flow velocity is 0.37 m/s  Front panel. It's made of toughened glass, which can not be caused negative effect by cleanliness and sterilization. Thickness of front panel is no less than 5mm  UV light. UV interlock provides additional safety by prohibiting UV light operation unless the front window is fully closed..

Page 7 (3m 41s)

[Audio] WORKING OF CABINET WORKING BIO SAFETY CABINET  HEPA Filter 99.999% efficient (0.3 microns). HEPA filter is constructed of pleated borosilicate glass fibers. HEPA Filter ( Standard for Manual Window Series) 99.999% efficient( 0.3 microns) HEPA filter is constructed of pleased borosilicate glass fibers. Filter Life Indicator. The pressure value is displayed on the control panel to show the life utility of the main filter and exhaust filter..

Page 8 (4m 26s)

[Audio] TECHNICAL SPECIFICATION Model No BTI- BSC – 3 2 2 Opening Height 250 MM SASH OPENING ULPA Filter ULPA FILTER WITH GREATER THAN 99.999% EFFICIENCY HEPA Filter EXHAUST HEPA FILTER WITH GREATER THAN 99.99[break]% FRONT WINDOW TOUGHENED GLASS GREATER THAN EQUAL TO 5 MM. NOISE LESS THAN 65 Decibel UV LAMP ONE FLUORESCENT LAMP 20 W * 1 LIGHT INTENSITY 900 LUX CONSUMPTION 800 WATT.